Two-Line Method

To implement the Two-Line Method, select Tools > Calibration Wizard to open the wizard window. Then select “Two-line method” on the Type drop-down menu. The “two line” calibration method removes the effect of the probe fixtures from a device under test and assumes the effect of fixture 1 is identical to fixture 2.

Note:

The two line method assumes the probe discontinuity can be represented as a simple shunt element. A larger probe will generate proportionally larger errors in calibration. Tolerance is at user discretion.

Note:

There can be only one propagating mode on the transmission line. If there are multiple propogating modes, then calibration may fail.

The required input is:

  1. Two Touchstone files (*.s2p), each of a thru-line of a given length plus the probe fixtures. The lines must be of different lengths.
  2. The imported Zo uses Touchstone file (*.s2p) of an ideal line with no fixture effects. This Touchstone file may come from measurement or simulation. If no file is specified, the wizard approximates the Zo.
  3. A Touchstone file (*.s2p) of the chosen device under test with fixture effects.

The output is: